| Alarm range (SM6): | 15 -2.0 m/s (30-400 fpm) | |
| Alarm range (12150): | 15 -10.0 m/s (30-2000fpm) | |
| Field set-up: | Single or 2 Point pushbutton calibration (Installer selectable) | |
| Accuracy: |
Face velocity accuracy / duct velocity accuracy +/- 10% |
|
| Airflow Sensor: | On board or remote sensor | |
| Low Air Alarm Delay: |
1 Low Air Alarm |
|
| Relay output: |
2 - Sash High and Night Setback |
|
| Comm. Port: |
RS485 – for monitor configuration only |
|
| Sash high indication: |
Yes, with separate plug-in connection with alarm repeat feature factory set to 5 minutes. (can be adjusted via laptop up to 30 minutes) |
|
| Night setback: |
Yes – Using a relay input |
|
| External alarm indication: | Not available | |
| Power Requirement: | Input—240VAC 50Hz, (UK) or 230VAC 50Hz (Euro) or 120VAC 60Hz (USA) Output—15VDC, 500ma | |
| Display - visual: |
LEDs: red, alarm; yellow, caution; green, normal |
|
| Alarm indication: | Red LED & audible alarm (to a certain sound spec) | |
| Horn silence: |
Yes (temporary or permanent) |
|
| Mounting: | Flush (or surface with optional mounting box) | |
| Monitor operating temp: |
55-86°F (13-30°C) |
|
| SM6 operating temp: | 15-25°C (Environmental limits) | |
| SM6 working temp: | ambient +25°C | |
| SM6 output mV: |
0-90 mV over working range (0-1.5m/sec). |
|
| Storage temperature: |
-40-150°F (-40-65°C) |
|
| Agency listings: | CE | |
| Hazardous area: | SM6 Suitable for Gas group llC Temp Class T6 (With Atex certified MTL Zener Barriers). (Remote sensor type only). |
| 5 | MK2 | UK | SM6 | 2M |
|---|---|---|---|---|
| Model | Style | Power Supply | Airflow Sensor Type | Sensor Cable |
| 5 | MK2 | US - 120v 60Hz Euro - 230v 50Hz UK - 240v 50Hz |
None (Built in sensor) SM6 12150 ILS |
2M 5M |
The AFA500 Airflow monitor is supplied complete with a plug in type power supply for connection to a single gang socket. The mains supply should be fused 3A.
A desktop style power supply can be supplied with flying mains lead for connection to a fused spur or terminals.
The standard AFA500 is supplied with a built in sensor. A remote type airflow sensor version is available and has a plug in RJ11 socket on the rear of the case to accept the sensor cable.
A vent it can be supplied for connection to the Fume Cupboard inner sidewall, this comprises of 1M of 25mm vent tube, a 25mm Bush and Coupling and box lid. See EVK on accessories page.
The standard length of a remote type sensor cable is 2M, a 5M version is also available. Custom lengths can be supplied on request.
The AFA500/MK2 can be supplied with a custom overlay (details on request).
The AFA500 airflow monitor range includes the following accessories:-
The AFA500 airflow monitor will operate with any of the following remote airflow sensors:-